Consumables
Gas Supply
Thermo Scientific
iCAP RQ Pre-Installation Requirements Guide (P/N BRE0009927, Revision A)
5-13
Pure Oxygen or Mixtures with Helium and/or Argon
For the analysis of organic solvents, the use of oxygen may be required.
The gas requirements are detailed in
.
NOTICE
A temperature of the spray chamber down to -10 °C will be
required for the analysis of organic solvents. This is controlled via the
Peltier cooler.
▲
Pure Nitrogen or Mixtures with Helium and/or Argon
For the analysis of organic solvents, the use of nitrogen may be required.
The gas requirements are detailed in
.
Methane Mixtures with Argon
NOTICE
The maximum content of methane is 4.5%.
▲
The gas requirements are detailed in
.
Table 5-12.
Additional gas requirements for oxygen
Gas Requirement
Specification
Purity
99.996% or better
Maximum Water Content
5 ppmv
Supply Rate
max. 1 L/min per channel
Pressure Regulator Output
max. 1 bar using a two-stage
pressure regulator (200 bar to
3 bar)
Table 5-13.
Additional gas requirements for nitrogen
Gas Requirement
Specification
Purity
99.996% or better
Maximum Water Content
5 ppmv
Supply Rate
max. 1 L/min per channel
Pressure Regulator Output
max. 1 bar using a two-stage
pressure regulator (200 bar to
3 bar)
Table 5-14.
Additional gas requirements for methane
Gas Requirement
Specification
Purity
99.996% or better
Maximum Water Content
5 ppmv