Line Power
Electrical Power Requirements
4-2
iCAP RQ Pre-Installation Requirements Guide (P/N BRE0009927, Revision A)
Thermo Scientific
Electrical Power Requirements
The iCAP RQ instrument is designed to operate at a nominal voltage of
200-240 V, 50/60 Hz. For details, see
Due to fluctuations, the temporary line voltage can vary between a
minimum of 180 V AC and a maximum of 264 V AC.
NOTICE
In areas with low line voltage (208 V in North America, 200 V
in Japan) during high use periods, voltage sags may occur which exceed
the save working voltage range of the instrument. In this case, you must
protect your instrument by using a suitable power conditioner or
uninterruptible power supply (UPS).
▲
The iCAP RQ instrument must have a separate “clean” line leading to a
main fuse to guarantee disturbance-free operation.
Power Connections
The electrical wall outlet for the main power of the iCAP RQ
instrument should be located at the wall near the intended location of
the instrument, ideally at the left side of the instrument.
The instrument is shipped with a standard IEC 60320 socket and will
be shipped with power cables suitable for your country. Depending on
your area, different types of plugs and cables may be required.
NOTICE
For more information on the correct cable kit, contact your
certified Thermo Fisher Scientific service representative.
▲
Power cables and connectors for the options are standard equipment
delivered by the manufacturers.
Table 4-1.
Basic power requirements
Specification
Value
Nominal Voltage
200-240 V AC, 50/60 Hz AC, single phase
Power
apparent power: 3000 VA;
effective power: 2200 W
Wire
Single-phase, 2-pole, 3-wire grounding
Fuse
Thermo Fisher Scientific recommends fusing
the wall receptacle with 15/16 A (Europe) and
20 A (North America, Japan).