iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision A)
5-5
Gas Supply iCAP RQ ICP-MS
The iCAP RQ instruments require argon gas to generate the inductively
coupled plasma and for controlling internal functions with the aid of
pneumatics. Additional gases might be required depending on the type
of analysis planned.
The iCAP RQ instruments use approximately 16 L/min (max.
24 L/min) of argon. It is essential that the gases be delivered with the
necessary pressure and purity. In case of questions, contact your certified
Thermo Fisher Scientific service representative.
Gas Supply Connections
shows the torque plate with the gas connections for the
iCAP Qnova Series ICP-MS 1-channel systems.
shows the torque plate with the gas connections for the
iCAP Qnova Series ICP-MS 2-channel systems.
NOTICE
All gas flow rates are given to standard conditions.
CAUTION
When operating the instrument with hazardous gas, an external
warning sensor must be installed in the vicinity of the instrument.
NOTICE
Note the different torque plates for iCAP RQ 1-channel (
and 2-channel systems (
Labeled Components: 1=additional gas 1 inlet, 2=additional gas 2 inlet,
3=argon gas inlet, 4=CCT inlet
Figure 5-4.
Gas connection at the left of iCAP RQ 1-channel system
1
2
3
4