Consumables
Cooling Water
5-4
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision A)
Thermo Scientific
Technical Data of Recirculating Chiller
The iCAP Qnova Series instruments are optionally delivered with the
Thermo Scientific ThermoFlex™ 2500 recirculating chiller with closed
circuit, cooled by a refrigerating device. A schematic of the
water-circulation including a filter is shown in
lists the technical specifications for the Thermo Scientific
ThermoFlex 2500 recirculation chiller, which is optionally shipped with
the instrument.
Figure 5-3.
Cooling circuit with recirculating water chiller
Table 5-2.
Chiller specifications
Chiller Component
Specification
Cooling Capacity
(at 20 °C ambient temperature)
2.2 kW (50 Hz); 2.5 kW (60 Hz)
Ambient Temperature
10 to 40 °C
Water Temperature
5 to 40 °C
Stability of Temperature
Regulation
±0.1 °C
Reservoir Volume
7.2 L
Flow Rate
Turbine Pump (60 Hz)
Turbine Pump (50 Hz)
13.3 L/min at 0.41 MPa (4.1 bar)
9.5 L/min at 0.41 MPa (4.1 bar)
Unit Dimensions (H × W × L)
73.6 cm × 43.6 cm × 67.3 cm
Unit Weight
79.6 kg
Instrument
Chiller
Water In
Water In
Water Out
Water Out
Filter
NOTICE
For information on the power supply of the chiller, see
.