Site Preparation
Dimensions and Weights
Thermo Scientific
iCAP Qnova Series ICP-MS Pre-Installation Requirements Guide (P/N BRE0006946, Revision B)
2-13
shows typical data system hardware components.
lists space requirements and weights of the typical data system
hardware components. The actual values depend upon your equipment.
The USB communication cables between the computer and the mass
spectrometer must be no longer than 3 m each. The workbench that
holds the data system must be located next to the workbench that holds
the mass spectrometer. For a connection length of more than 3 m, an
active hub must be placed between the instrument and the computer.
The length of the USB cable between the iCAP Qnova Series
instrument and the hub is limited to 3 m, whereas between the hub and
the computer a cable length of 5 m can be used. This hub must be used
exclusively for the control of the mass spectrometer.
Figure 2-12.
Typical data system workbench
Table 2-2.
Typical data system space and load requirements
Module
Height [cm]
Width [cm]
Length [cm]
Weight [kg]
Monitor
36
41
18
6
Minitower
computer
48
18
43
14
monitor
computer
wall outlet for data system components
120 or 230 V AC
communication
cables USB and
Ethernet