System Setup
Placing the System
Thermo Fisher Scientific
iCAP Q Operating Manual (P/N 1288090, Revision B)
5-5
Moving the Forepump
For easy and effortless transportation of the forevacuum pump, Thermo
Fisher Scientific provides a mobile base (drip pan with casters) with the
iCAP Q mass spectrometer.
If lifting of the forevacuum pump becomes necessary, only a suitable
lifting device shall be used to lift the pump at the lifting lug due to the
forevacuum pump weight of about 43 kg.
Should people lift the pump, at least two persons should lift it. Pay
attention to the persons' position when lifting! Straight back, flexed
knees, and adapted back large belts, etc. are recommended.
When moving the forepump, pay attention to the following guidelines:
•
1 person stands on the left side, one hand under the oil casing and
one hand under the pump stator.
•
1 person stands on the right side, one hand under the oil casing and
one hand under the motor fan cover.
•
Attention to your hands when setting down the pump into the drip
pan to avoid squashing!
NOTICE
Do not lift the pump at the frequency converter or the
cables!
▲
Workbench for Instrument
The workbench for the MS system must be capable of supporting the
weight of the mass spectrometer(148 kg) plus the weight of any option
(autosampler, liquid chromatograph, for example) and stand in a secure
and level position.
CAUTION
!
Heavy Instrument.
Danger of injury. This instrument is too heavy for
one person alone to handle safely. Lifting and moving the instrument
requires the effort of at least
four
persons. Thermo Fisher Scientific
recommends using a
pallet jack to lift
the mass spectrometer onto the
workbench.
!
A lifting device is the only officially recommended way of handling the
pump!
!
Only workbenches with four legs provide sufficient stability for the
iCAP Q mass spectrometer. The workbench must be dry and clean.