System Description
Gas Supply
Thermo Fisher Scientific
iCAP Q Operating Manual (P/N 1288090, Revision B)
3-17
Additional Gas
Depending on the application, a variety of additional gases may be
employed. The use of mass flow controllers is required when working
with additional gases.
Oxygen
For the analysis of organic solvents the use of oxygen may be required.
The gas requirements are detailed in
.
NOTICE
Temperature down to -10 °C will be required for the analysis
of organic solvents. This is controlled via the Peltier cooler.
▲
Helium
For laser ablation analysis, helium gas may be needed. The gas
requirements are detailed in
Table 3-5.
Additional gas requirements for oxygen
Gas Requirements
Specification
Purity
99.996%
Maximum water content
< 5 ppmv
Supply rate
max. 1 L/min
Pressure regulator output
0.6 MPa (6 bar)
Table 3-6.
Additional gas requirements for helium
Gas Requirements
Specification
Purity
99.996%
Maximum water content
< 5 ppmv
Supply rate
max. 1 L/min
Pressure regulator output
0.6 MPa (6 bar)