Additional Safety Information
Fore Vacuum Pump
9-2
iCAP Q Operating Manual (P/N 1288090, Revision B)
Thermo Fisher Scientific
Fore Vacuum Pump
Failure to observe the following precautions could result in serious
personal injury!
Sogevac™ pumps are not designed:
•
For pumping of dusty, aggressive, corrosive, flammable or explosive
gases or gases mixtures,
•
For pumping of oxygen or other highly reactive gases with a greater
concentration than atmospheric concentration (>20%),
•
For working in flammable, explosive or dusty environment.
For all these cases, special materials must be used. In case of doubt,
please contact Oerlikon Leybold Vacuum. See also the limits of use
indicated in the CE declaration of conformity.
Pump must not be operated above 2000 m sea level.
WARNING
!
High Voltage.
When touching parts at high electric voltages, there is
the risk of suffering severe injuries by an electric shock! Covers marked
with this symbol must only be opened by trained electricians after
having reliably deenergized (lockout/tagout) the equipment.
Always operate the pump with a properly connected protective earth
conductor and make sure that the motor & FC connection box are
closed.
Observe the manufacturer’s information and operating instructions for
the respective frequency converter (enclosed).
Use only the OLV frequency converter for the pumps equipped with
one.
After having made changes to the wiring, check the motor’s direction of
rotation.
WARNING
!
High Voltage.
Even if the pump is not operating, there is the risk of
suffering severe injuries by an electric shock! Even after complete power
disconnection, live voltage is present in the frequency converter. Allow
a minimum of 10 minutes for a complete capacitor discharge.
WARNING
!
Explosion Hazard.
The standard pump is not suitable for installation
in explosion hazard ATEX areas. Contact the pump manufacturer if
you are planning such an application. Before installing the pump, you
must reliably disconnect it from the electrical power supply and prevent
the pump form running up inadvertently.