Thermo Scientific HM525 NX Operator Guide
15
Technical Specifications
Mechanical Specifications
Depth
760 mm (29.9 in)
Width
640 mm (25.19 in) w/o handwheel
Height Range
1150 mm ( 45.3 in)
Maximum Weight
143 kg (315.2 lb)
Electrical Specifications
Power Requirements
Dedicated line recommended.
100V, 50/60Hz, 7A, +/-10%, 110-120V, 60Hz, 7A, +/-10%
220-230V, 50/60Hz, 4A, +/-10%
BTU
1020BTU/hour
Environmental Specifications
For Indoor Use Only
Temperature (Operating Limits)
+5°C to +35°C (+41°F to +95°F)
(at a max. Relative humidity of 60%)
Temperature
(Recommended Operation)
+15°C to +30°C (+59°F to +86°F)
Note
Performance may deteriorate when operated outside of this
range.
Temperature
(Transport and Storage)
-20°C to +50°C (-4°F to 122°F)
Relative Humidity
Max. 60% RH up to 35°C
Altitude
Up to 2000 m (6,500 ft)
Pollution Degree
2
Over Voltage Category
II
Floor Loading Requirements
230 kg/m² (507 lb/ft²)
Protection Class
I
Sound Emission
45 dB (A) - measured with 1m distance to the instrument.
Refrigerant
R404a, filling amount: 100...120V 175g 220...240V 175g