3
Removing and Installing the HESI-II Probe
Removing the HESI-II Probe
Thermo Scientific
HESI-II Probe User Guide
11
4. Unplug the vaporizer cable from the HESI-II probe vaporizer cable socket (see
and
Figure 9.
Top view of the API source housing and the HESI-II probe
5. Disconnect the auxiliary gas fitting (green) from the auxiliary gas inlet (A) on the probe
(see
6. Disconnect the sheath gas fitting (blue) from the sheath gas inlet (S) on the probe (see
Figure 10.
Top view showing the sheath and auxiliary gas lines connected to the probe
A
S
Vaporizer cable
(connected to
the probe)
8 kV cable
(disconnected)
Probe vaporizer
cable socket
A
S
Sheath gas
line
Auxiliary gas
line
Vaporizer cable
(disconnected)
8 kV cable
(disconnected)