4
Operation
Centrifugation
4-10
Multifuge X1 / X1R
Thermo Scientific
The display restores the current values. Return to step 1.
Centrifugation
Once the rotor has been properly installed, the main switch has been turned on and the lid has been
closed, you can start centrifuging.
Starting centrifuge program
Press the START key
on the control panel. The centrifuge accelerates to the pre-set speed with the
time display active.
If the speed setting is higher than the maximum permissible speed or RCF-value for the particular
rotor, then the display will show the message CHECK SET SPEED. The centrifuge decelerates. Please
enter a new value.
You cannot open the lid as long as the centrifuge is running.
Imbalance indicator
If a load is imbalanced, this will be indicated at speed higher than approx. 300 rpm by the message
"Imbalanced load".
The run will terminate.
Check the loading and start the centrifuge once again. See the information on proper loading in the
rotor instruction manual. For information on troubleshooting, see section
“When to contact Customer
Service”
on
page 7-5
.
Stopping the centrifugation program
With preset running time
Usually the running time is preset and you only have to wait until the centrifuge stops automatically
when the preset time limit expires.
As soon as the speed drops to zero, the message END will appear in the display. By pressing the OPEN
key
, you can open the lid and remove the centrifuge material.
You can also stop the centrifuging program manually at any time by pressing the STOP key
.
Continuous operation