Concentration Measurements—Standard Curve Application
Measuring Samples
86
GENESYS 10S Vis User Guide
Thermo Scientific
If a 6-Position Cell Holder is installed, press the cell position buttons to reposition the cell
holder and measure the rest of the standards manually.
When all of the standards have been measured, the Standards screen (see
) shows the absorbance of each standard, along with the
slope, intercept and correlation coefficient of the standard curve.
Using the Standards Screen
Here is an example of the Standards screen showing measurement results:
You can use this screen for:
• Displaying a graph of the standard curve data (press
View Graph
)
•
(press
Save Test
)
•
(press
Edit Standards
)
•
(press
Run Test
)
Measuring Samples
Y
To measure samples automatically using the calibration curve (using Auto 6 or Auto
3)
1. Press
Run Test
.
2. Install the blank and samples.
3. Press
Enter
.
The Standard Curve screen shows the absorbance and concentration of each sample.
To switch between tabular and graphical displays, press
View Graph/Tabular
.
Содержание GENESYS 10S Vis
Страница 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Страница 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 22: ...2 Setting Up the Instrument Setting Up the Internal Printer 14 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 56: ...11 Abs and T Measurements Advanced A T C Taking Measurements 48 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 64: ...13 Concentration Measurements Advanced A T C Measuring Samples 56 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 106: ...20 Kinetics Rescaling and Recalculating Tabular Kinetics Results 98 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...