21
Performance Verification
Photometric Accuracy
104
GENESYS 10S Vis User Guide
Thermo Scientific
Photometric Accuracy
This test measures the absorbance (or %T) of a set of standards and compares the results with
specified tolerances. The wavelength absorbencies and tolerances are preset, but you should
change them to the values on the calibration certificate included with your standards.
When running the Photometric Accuracy test, remember:
• You need Photometric Accuracy standards calibrated to known absorbance values at
specified wavelengths.
• Measure the standards in the order they are listed on the test screen.
• You can use one to five standards.
Y
To display the Photometric Accuracy screen
1. Highlight
Photometric Accuracy
.
2. Press
Enter
.
Selecting the Mode
To switch between absorbance and %Transmittance, press
Change to %T
(or
Change to
ABS
) until the appropriate mode appears.
Adding Standards
You will need to set three values whenever you add a standard: the wavelength, the absorbance
(or %Transmittance) and the tolerance value.
Note
You can display the tolerances for this test in either absorbance or %Transmittance.
Содержание GENESYS 10S Vis
Страница 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Страница 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 22: ...2 Setting Up the Instrument Setting Up the Internal Printer 14 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 56: ...11 Abs and T Measurements Advanced A T C Taking Measurements 48 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 64: ...13 Concentration Measurements Advanced A T C Measuring Samples 56 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 106: ...20 Kinetics Rescaling and Recalculating Tabular Kinetics Results 98 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Страница 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...