Chapter 15
Maintenance and Troubleshooting
Maintaining a PPKD Injector
276
Maintaining a PPKD Injector
The Purged Packed Column injector and its interface module will normally be
serviced by Thermo Fisher Scientific authorized technical personnel.
In order to operate at peak performances, the injector requires periodic
maintenance from the user. This maintenance includes:
•
the replacement of the septum
•
the cleaning or replacement of the liner.
When replacing the septum
The septum must be replaced at least every 200 injections, or every time a
problem occurs related to a septum damage or wear.
It is good practice to replace the septum every time the liner is replaced.
When replacing the liner
The liner must be replaced periodically, depending on the number of injections
performed and on the characteristics of the samples injected. The most common
symptom indicating that the liner should be replaced is the appearance of tailing
peaks in the chromatogram, particularly for polar compounds.
You can replace the liner with a new one or clean the liner and reinstall it.
Every time the liner is replaced, it is a good practice to replace the septum too.
The following Operating sequences help you to maintain properly the PPKD
injector:
•
•
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...