OPERATING SEQUENCE
Replace the S/SL Inlet Vent Line Filter
This carbon filter is mounted on the DPFC module located on the rear of the GC.
To replace the filter operate as follows:
WARNING! Before operating the FOCUS GC MUST BE SWITCHED OFF.
Remove the FOCUS GC Top Cover
The top cover is fixed on the main unit with two front screws and two clips.
1. Open the GC oven door.
2. Unscrew the two front screws, as shown in Figure 15-6, which remain on the
main unit.
3. Remove manually the top cover pulling it up from the front, the two clips will
be automatically unlocked.
Figure 15-6. Top Aluminium Cover Removal (1)
Remove the FOCUS GC Rear Panel
4. Unscrew the four Allen fixing screws, as shown in A of Figure 15-7 .
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...