Chapter 15
Maintenance and Troubleshooting
Maintaining a SSL Injector
270
Figure 15-3. Replace a Broken Liner
c.
With the aid of a pipe cleaner, remove the possible glass fragments from
the vaporization chamber.
d. Reinsert the silver seal and the terminal fitting, and then tighten the nut
that retains the terminal fitting. See F of Figure 15-3.
e.
Insert the analytical column with its ferrule into the bottom of the injector
in its previous position, then tighten the M4 retaining nut to hold the
column in place. See G of Figure 15-3.
Replacing the Liner
8. Holding the new (or cleaned) liner with tweezers, place a graphite seal over
the liner, making sure to leave a distance of about 10 mm between the seal and
the liner end.
9. Using tweezers, insert the liner into the injector and push it gently towards the
bottom fitting. See D of Figure 15-4.
D
E
F
G
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...