Chapter 8
FID Menu
Flame Ionization Detector (FID)
165
Flame Out Conditions
When the flame is accidentally extinguished, either permanently because of
exhausted fuel gas supplies or temporarily, the
Flame Out
message is displayed
in the menu and a message is recorded in the Run Log.
If the
Retry
function is turned
On
, the system will attempt to re-ignite the flame
up to three times.
Signal pA
0 - 1000.00 pA
Not editable
This parameter shows the collector current in pico
amperes (standing current level). The displayed value is
also used to indicate the flame status. If the value is very
low (such as 0.3 pA), the flame is off. When the value
displayed is greater than the
Ignition threshol
d,
the flame is on.
Ign. thresh
0 - 999 pA
The FID produces a small signal current when lit. This
parameter defines the flame on condition. The FOCUS
GC uses this value to determine flame status (on or off)
and control automatic re-ignition. If
Flameout retry
is
O
n,
the flame will re-ignite if the signal drops below
this value. This line doesn’t appear at detector signal
range 2 and 3. (Threshold assumes 0 pA).
Flameout
retry
On – Off
This indicates re-ignition status. Set On to program
when the flame re-ignition should be attempted. Refer to
for more information. This line
doesn’t appear at detector signal range 2 and 3.
H2
Air
On - Off
These indicate the hydrogen and air flow supplied to the
detector. You can only turn the flows on and off. Set On
to turn on the gas flows. Set Off to turn the flows off.
Make-up
On - Off
This indicates the make-up gas used with the FID. You
can only turn the flows On and Off. Set On to turn on
the gas flows. Set Off to turn off the flow.
Table 8-1. FID Parameters (Continued)
Parameter
Range
Description
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...