Chapter 7
Column
Installation
140
Connecting a Packed Column to the S/SL Injector
Materials required:
•
liner cap removal tool
•
retaining nut
•
ferrules
•
metal adapter
1. Remove the S/SL Injector Top Components.
With reference to Figure 7-6 proceed as follows:
Figure 7-6. Removing the S/SL Injector Top Components
a.
Unscrew the injector cap.
b. Remove the septum holder with septum. then the septum support.
c.
Remove the liner cap by using the tool provided.
d. Use tweezers to remove the liner with the graphite seal.
1. Injector Cap
2. Septum Holder
3. Septum
4. Septum Support
5. Liner Cap Removing Tool
6. Liner Cap
7. Liner
8. Graphite Seal
1
2
3
4
5
6
7
8
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...