background image

3

TIP EJECTION (Fig.3)

1. To remove the tip and plunger from the pipette simply

press the operating button all the way down.

2. Then release the operating button.
3. This action simutaneously ejects the tip and the

plunger.

PIPETTING TECHNIQUE

Figure 6
A = Ready Position

B = Pipetting Stop
C = Tip Ejection

Ensure that clean tips are firmly pushed on to the tip cones

of the pipette.
The operation of the pipette is controlled by the operating

button. For the best possible accuracy:
Operate the button slowly and with steady speed at all

times, particularly when working with liquids having high
viscosity. Never let the operating button snap back. When

working with serum or biological fluids prime the tips with
the liquid before pipetting. This is done by filling and

emptying the tips.
Hold the pipette vertically (10 degrees maximum from

vertical) during the liquid intake. For maximum hand
control, hold the pipette tightly in the hand with the

"Grippy" resting on the index finger.
Fill a clean reagent reservoir with the liquid to be

dispensed.
1. Depress the operating button to the pipetting stop.
2. Dip the tip just under the surface of the liquid in the

reservoir (2-3 mm) and slowly release the operating
button. Withdraw the tip from the liquid touching them

against the edge the reservoir to remove excess liquid.

3. Deliver the liquid by gently depressing the operating

button to the pipetting stop.

4. Release the operating button to the ready position.
If necessary, change the tips and continue with the
pipetting.

Содержание FINNPIPETTE PDP

Страница 1: ...tems GRIPPY HANDLE DIGITAL DISPLAY FINNTIPPDP OPERATING BUTTON FINNPIPETTE PDP INSTRUCTIONSFORUSE Thermo Electron Oy P O Box 100 Fin 01621 Vantaa Finland Tel 358 9 329100 fax 358 9 32910414 www thermo com finnpipette ...

Страница 2: ...tus Corporation DESCRIPTIONOFFINNTIPPDP Fig 4 FINNPIPETTE PDP tips are supplied in racks of 96 or bags of 960 The product code numbers are as follows Cat No Volume Qty 9403070 0 5 25µl 960 bag 9403080 20 200µl 960 bag PIPETTEOPERATION SETTINGTHEDELIVERYVOLUME Fig 2 A The delivery volume is set using the operating button on the top of the pipette To increase the delivery volume turn the operating b...

Страница 3: ...h serum or biological fluids prime the tips with the liquid before pipetting This is done by filling and emptying the tips Hold the pipette vertically 10 degrees maximum from vertical during the liquid intake For maximum hand control hold the pipette tightly in the hand with the Grippy resting on the index finger Fill a clean reagent reservoir with the liquid to be dispensed 1 Depress the operatin...

Страница 4: ...r 2 Pull away the tip cone cover 3 Remove the tip cone by turning it counterclockwise with the service tool 4 Pull out the piston 5 Clean the piston and the piston spring with a dry napless cloth 6 Grease the cleaned parts with lubricant provided in the package 7 Reassemble the parts CALIBRATION All Finnpipettes are factory calibrated and adjusted to give the volumes as specified with distilled or...

Страница 5: ...ipette is correct Procedure 1 Do 10 pipettings with the lower volume 2 Do 10 pipettings with the maximum volume 3 Calculate the accuracy A and precision cv of both series 4 Compare the results to the limits in the Table 1 If the results are in the limits of Table 1 then the calibration of the pipette is correct Otherwise the pipette must be adjusted and checked again Range Volume Accuracy Precisio...

Страница 6: ...g the weight of the water to volume at test temperature and pressure A typical value is 1 0032 µl mg at 22 C and 95 kPa See the conversion table on page 8 Accuracy systematic error Accuracy is the difference between the dispensed volume and the selected volume of a pipette A V V0 A accuracy V mean volume V0 nominal volume Accuracy can be expressed as a relative value A 100 x A V0 Precision random ...

Страница 7: ...equire recalibration Solution Attach firmly Clean tip cones thoroughly and attach new tips Follow instructions carefully Attach firmly Recalibrate according to instructions Recalibrate with the liquids in question CAUTION The Finnpipette is designed to allow easy in lab service If you however want to send the pipette to us or to our local representative for service please enclose a list of any inf...

Страница 8: ...22 1 0031 1 0031 1 0032 1 0032 1 0033 1 0033 22 5 1 0032 1 0032 1 0033 1 0033 1 0034 1 0035 23 1 0033 1 0033 1 0034 1 0035 1 0035 1 0036 23 5 1 0034 1 0035 1 0035 1 0036 1 0036 1 0037 24 1 0035 1 0036 1 0036 1 0037 1 0038 1 0038 24 5 1 0037 1 0037 1 0038 1 0038 1 0039 1 0039 25 1 0038 1 0038 1 0039 1 0039 1 0040 1 0041 25 5 1 0039 1 0040 1 0040 1 0041 1 0041 1 0042 26 1 0040 1 0041 1 0042 1 0042 1...

Страница 9: ...9 2 5 0 Labsystems 1 2 3 3 2 5 0 Labsystems 5 25 µl 200 µl 4 ...

Страница 10: ...10 absystems 6 A C B 1 2 3 4 7 Pipetting Tip Ejection ...

Страница 11: ...500 3 Handle Back 10588030 10588030 4 Limiter 1092990 1092990 5 Handle 2204030 2204040 6 Locking Pieces 1058180 1058180 7 Plunger 1106230 1106300 8 Oper Spring 1130720 1130720 9 Ring 1006500 10 Tip Cone 1058790 10588000 11 Tip Cone Cover 10588011 10588011 SPAREPARTS 1 2 4 5 3 6 7 8 9 10 11 ...

Страница 12: ...ns are subject to change without prior notice Finpipette and Finntip are registered trademarks of Thermo Electron Oy Thermo Electron Oy P O Box 100 Fin 01621 Vantaa Finland Tel 358 9 329100 fax 358 9 32910414 www thermo com finnpipette ...

Отзывы: