Setting Up the Xcalibur Instrument Configuration
Specifying the Reagent Ion Source for the MS Detector
10
MS/ETD System Getting Started Guide
Thermo Scientific
Specifying the Reagent Ion Source for the MS Detector
Specify the ETD module as the reagent ion source for the LTQ XL or LTQ Velos MS
detector.
To specify the reagent ion source
1. In the Thermo Foundation Instrument Configuration window, in the Configured
Devices list, double-click the
LTQ XL MS
or
LTQ Velos MS
icon to open the
Configuration dialog box (
).
Figure 4.
LTQ XL Configuration dialog box
2. In the left pane, select
Reagent Ion Source
.
The reagent ion source configuration appears (for the LTQ Velos MS detector, see
).
3. Select the
Reagent Ion Source Configured
check box.
You must manually select this check box for MS/ETD systems.