background image

 Page 

7  

Enter a time parameter as described above for setting a voltage or current setpoint.  Enter a time 
duration of 0-99 hours, 59 minutes.  A time duration of zero effectively disables timed run mode. 

When running a timed run, the time parameter displays the time remaining in the run.  When 
running a non-timed run, the time parameter displays the elapsed time during the run 

Automatic Power-fail Restart 

Automatic power-fail restart capability allows a timed run which is interrupted by loss of AC 
power to be restarted automatically, so that the total time programmed for the time parameter will 
be met.  

NOTE:

 Loss of AC power includes switching off the unit using the power switch located 

at the rear of the unit.  

When performing timed runs with power-fail restart enabled, always use 

the front panel controls to stop a run in progress. 

To enable power-fail restart, press the top function key when in the operating mode screen 
(constant voltage or current).  The state of power-fail restart will toggle from off to on or on to off 
with each keypress. 

When AC power is restored during a timed run in which power-fail restart is enabled, the display 
will show "POWER-FAIL RESTART" for approximately 10 seconds indicating power-fail restart 
pending.  During this time the output is not energized, to allow safely stopping the run (by 
pressing Run).  After the power-fail restart pending delay is completed, the power supply output 
will ramp up to the setpoint value and the time will continue from that point.  Any number of 
power interruptions can occur during the completion of a single run. 

Cleaning 

Before cleaning the unit, be sure to always turn off AC power using the switch at the rear of the 
unit, and then disconnecting the AC line cord.  The front of the unit is sealed and can be wiped 
clean with any mild detergent solution.  Avoid harsh cleaners or agents as they may deteriorate the 
surface of the tactile membrane keys. 

Содержание EC200XL

Страница 1: ...LTM1595X2 8 28 07 2007 All rights reserved Printed in U S A EC200XL Electrophoresis Power Supply ...

Страница 2: ...on 3 Unpacking the Power Supply 3 Specifications 4 Getting Started 5 Using the Power Supply 5 Constant Voltage Operations 6 Constant Current Operations 6 Timed Operations 6 Automatic Power fail Restart 7 Cleaning 7 Troubleshooting and Error Indications 8 EC200 Warranty Statement 9 Compliance 9 Replacement Parts 9 ...

Страница 3: ...ated apparatus for any signs of wear or damage before using with this equipment Be sure to use only electrophoresis equipment that is suitably rated for the voltage and current capabilities of the EC200 power supply The output of the EC200 power supply is intended for connection to electrically isolated electrophoresis apparatus only Use only with electrically isolated electrophoresis apparatus wi...

Страница 4: ... If this non constant limit is reached the power supply will automatically switch control modes from constant voltage to constant current or vice versa In this way the EC200 protects your electrophoresis cells from damaging over power conditions The EC200 power supply also provides for timed operation in either voltage or current modes and allows an automatic completion in the event of a power los...

Страница 5: ...0 60Hz 250VA Max Environmental Operating temperature 0 40 C 0 95 R H non condensing Altitude 2000m Overvoltage category II IEC664 Pollution degree 2 IEC664 DC output Power 5 200VDC 200W Max 0 01 2 00A 200W Max Ripple 1 Drift 1 after 30 minute warm up ...

Страница 6: ...ate and show the values of the last saved run setup The EC200 preserves the run settings each time you start a run One set of conditions is saved for each mode constant voltage or constant current The setpoint value limit parameter time duration and power fail restart settings are all saved This allows convenient setup for repetitive runs To change the control mode press the Exit key From the resu...

Страница 7: ...et to the power supply maximum value of 2 00 A The EC200 will not accept a current limit outside of the allowed range In addition the EC200 will never exceed the maximum power output specification of 200W When starting a run the power supply calculates a maximum limit parameter above which the power supply maximum power output rating would be exceeded If this value is lower than the limit setpoint...

Страница 8: ... restart press the top function key when in the operating mode screen constant voltage or current The state of power fail restart will toggle from off to on or on to off with each keypress When AC power is restored during a timed run in which power fail restart is enabled the display will show POWER FAIL RESTART for approximately 10 seconds indicating power fail restart pending During this time th...

Страница 9: ...ctions 02 Step load current change The EC200 has detected a large change in load current Check your setup and connections 04 Loss of AC power PF disabled and not completed The EC200 detected loss of AC power the run was not completed since PF was not enabled 05 Loss of AC power run not timed The EC200 detected loss of AC power 06 99 The EC200 detected an internal error condition Contact technical ...

Страница 10: ...chaser s exclusive remedy for any claim arising out of the purchase of any product listed above is repair replacement or credit as described above where applicable In no event 1 shall the cost of the exclusive remedy exceed the purchase price 2 shall Thermo be liable for any special indirect incidental consequential or exemplary damages howsoever arising even if Thermo has been advised of the poss...

Отзывы: