Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
31
2
Analytical Principles
This chapter describes the analytical techniques with the correlated pneumatic circuits used for the
configurations of the EA IsoLink IRMS System for CNSOH.
About Your System
The Elemental Analyzer is equipped with an automatic switch valve (ASV) between the two modes of
high temperature conversion
(
H
and
O
isotope analysis), and combustion (
N
,
C
, and/or
S
isotope
analysis).
The switch also involves the transfer of the electronic autosampler start signal to either side.
The EA IsoLink IRMS System for CNSOH is configured with an optimized reactor and separation
column for triple analysis of
N
,
C
and,
S
isotopes, and
Sulfinert®
capillary for minimized water
adsorption, and reduced nitrogen background.
The
NCS
separation column is operated in a ramped GC oven attached to the back panel of the Flash
IRMS EA. The molecular sieve for high temperature conversion analysis is installed in the isothermal
GC oven of the EA.
Two autosamplers on top and full software control automate the analysis of five isotopes with two
sample drops.
The additional bottom feed connector for the pyrolysis side extends the sample throughput to more
than 400 solid samples without maintenance.
This section describes the new features and software setup of the EA IsoLink IRMS System for
CNSOH, and shall be seen in addition to the existing:
•
ConFlo IV Operating Manual Rev. B
It is recommended to read the above manuals in addition to this guide.
The following analytical setups are possible: