Getting Familiar with your EA IsoLink IRMS System for CNSOH
Autosamplers
26
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Autosamplers
The EA IsoLink IRMS System for CNSOH can be configured with the following autosamplers:
MAS
Plus
for solid samples, and
AI 1310/AS 1310
for liquid samples.
MAS Plus Autosampler for Solid Samples
It is the standard autosampler for solid samples mounted directly on the connecting fitting of the
concerned channel provided with the proper reactor tube. See
.
Figure 26.
MAS Plus Autosampler for Solid Samples
Table 11.
Status LED Description
LED
Meaning
Power On
When lit, the instrument is powered on.
Ready
When lit, the instrument is ready to run analyses.
Run
When lit, an analysis is in progress.
Standby
When lit, the instrument is in standby condition.
During this condition, helium gas flows are decreased to 10 mL/min, and
the furnace temperatures reduced to 50% of the set value.
Wake Up
When lit, the instrument has been programmed for a timed automatic
startup (Ready Condition).
Safety Cutoff
When lit, the instrument is in safety shut-off. Gas flows are stopped,
furnaces are switched off and TCD is switched off.