Running the Flash IRMS as Stand-alone Instrument
Adjusting the Detector Signal Level
148
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Figure 156.
Leak Test Status Window
a.
Click
Start
to begin the leak test. The system automatically selects
Carrier
and
Reference gas
outlet closed
check boxes. The system will ask if the zero of flow controllers has to be
calibrated; reply
Yes
to this question. This operation will take place in less than one minute.
b. Gas outlets are closed by the solenoid valves. Wait for some time (see
Leak test time)
,
according to the instrument configuration, to let the gas circuit reach the equilibrium pressure.
From the values of
Carrier Flow
and
Reference Flow
must be within
0
and
3 mL/min
.
Higher values indicate that the system is not leak-free.
c.
Click
Stop
and
Done
for ending the leak test and restoring the flow operating values.
Adjusting the Detector Signal Level
To adjust the level of the TCD detector’s signal, follow the instructions reported in the operating
sequence
“To adjust the level of the detector signal”
.
To adjust the level of the detector signal
1. In Main Menu select
View
|
View Elemental Analyzer Status
,
or just click the icon
, then
select the
Detector
.
Figure 157.
Detector Status Tab
2. In the field
TCD
, click
Auto-Adjust Level at 1000
μ
V
. At the end of the operation, the value
1000 is set representing the analysis starting point.
Note
Leaks in the system are generally due to incorrect closure of the reactors and filters
locking nuts. Rarely, leaks may be due to the autosampler.