![Thermo Scientific EA IsoLink IRMS System Скачать руководство пользователя страница 139](http://html.mh-extra.com/html/thermo-scientific/ea-isolink-irms-system/ea-isolink-irms-system_operating-manual_1105021139.webp)
6
Using EA IsoLink IRMS System for CNSOH
Create a Gas Configuration for a Sulfur Measurement
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
111
4. The gases needed (He, O
2
) are available and have been connected to the corresponding positions.
5. Leak-check all connections outside of ConFlo by brushing all fittings carefully with soapsuds.
6. The exhaust tube is connected to the ConFlo and ventilation is working.
7. SO
2
is connected to the ConFlo.
8. The pressure regulator for SO
2
is heated.
9. The IRMS has been calibrated for SO
2
measurements.
10. The Elemental Analyzer is switched on.
11. Adjust the gas pressure of Elemental Analyzer and ConFlo. Take the following settings as a
guideline.
•
“Setting for Flash Elemental Analyzer”
Setting for Flash Elemental Analyzer
12. Increase reactor temperature to 900 °C. After 10 minutes at this temperature, increase to 1020 °C.
Alternatively, use the HeatUp function in the context menu of the Flash IRMS peripheral
visualization
13. After 10 minutes at this temperature, increase to 1020 °C. Increase column temperature to 100 °C.
14. It is recommended to keep the reactor over night at end temperature to stabilize the conditions.
Create a Gas Configuration for a Sulfur Measurement
A Gas Configuration determines a combination of masses, which are collected in the cups, for
evaluation of ratios and eventually
δ
values. The Gas Configuration is specific for the particular gas and
is combined with a magnet field value taken from the mass calibration of your IRMS. The ratio groups
determine the reported ratios of predefined masses.
Table 23.
Setting for Flash Elemental Analyzer in Eager
Smart
software
Carrier
180 mL/min
Oxygen
25 mL/min
Reference
20 mL/min
Cycle run time
1000 s
Sampling delay
10 s
Oxygen Injection End
3 s (or higher, depending on sample size and sample type)
Note
Teflon® tubings are not absolutely tight against atmosphere. Therefore, the backgrounds of
argon and nitrogen are higher than those of carbon and nitrogen measurements. Nevertheless, take
care of leaks.
Note
When using reduced amplification (i.e. resistor of 1×10
10
Ohm on cup for mass 66), the
background values will be three times higher.
Note
Background values may vary depending on sensitivity and focus settings.