Installing AI 1310/AS 1310 Autosampler
Installing the AI 1310/AS 1310 on the Flash IRMS Elemental Analyzer
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
81
3. Accommodate the syringe body into its seat paying attention to insert the flange and the head of
the syringe plunger simultaneously into their relevant guides.
4. Turn the lock knob by approximately 180° clockwise to lock the syringe. See
Figure 83.
Syringe Installation (2)
5. Close the safety door of the rotating turret.
Electrical Connections
To perform electrical connections
1. By using the cable provided, connect the 9-pin female connector marked
RS232
located on the
sampling unit back side to a 9-pin serial port connector (COM) of the PC.
2. Plug in the tampered connector provided into the 6-pin female connector marked
GC
located on
the sampling unit back side.
3.
Only for AS 1310
, connect the 15-pin female connector of the cable, coming from the support
plate of the 105-position sample tray, to the connector marked
TRAY
located on the back side of
the sampling unit.
Lock Knob