Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
Thermo Scientific
DSQ II Hardware Manual
75
Replacing the Calibration Gas Flow Module
To replace the calibration gas flow module, you will need:
• New Calibration Gas Flow Module (PN 119246-0001)
• FC-43 calibration compound (PN 50010-30059)
• Syringe
• Wrench, Allen, 3 mm
• Wrench, open-ended, 7/16-in.
• Wrench, open-ended, 9/16-in.
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Remove the left cover, as described in "
Removing the Left Cover" on page 29
4. Turn off the CI reagent gas supply (if present).
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
WARNING - FLAMMABLE GAS HAZARD.
CI reagent gases are often flammable or
corrosive. You must vent the gas to a fume hood or other suitable exhaust. The CI reagent
gas supply line must be leak tight.
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...