Troubleshooting
This chapter will enable you to identify any issues related to operating the DSQ II. Once you
understand the symptom, cause, and solution to the issue, use
to find out how to address it. This chapter is organized according to the type of
symptoms you may experience when using the DSQ II. If you experience issues that are not
addressed in this chapter, contact Technical Support.
Using Diagnostics
If your DSQ II is running poorly or suspiciously,
DSQ II Tune
diagnostics will most likely
detect the problem and recommend a solution. Diagnostics tests electronic circuits and
reports whether the circuits pass or fail.
However, problems in sensitivity due to misalignment, dirty components, or improper tuning
are not detected by diagnostics.
Before running diagnostics, refer to the flowchart
to determine solutions outside of
DSQ II Tune
diagnostics.
Contents
•
•
•
“Filament and Lens Control Symptoms”
•
•
•
•
•
•
•
•
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...