Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
Thermo Scientific
DSQ II Hardware Manual
113
Figure 68.
Removing the Transfer Line
a.
Disconnect the transfer line cable from the left side of the GC.
b. Loosen the transfer line gas fitting.
c.
Remove the three screws holding the transfer line in place.
d. Remove the transfer line.
6. Install the new transfer line (PN 119640-1050).
a.
Insert the new transfer line.
b. Reattach the screws that hold the transfer line in place.
c.
Tighten the transfer line gas fitting.
d. Reconnect the transfer line cable to the left side of the GC.
7. Reattach the right, top, and front covers of the DSQ II.
8. Restart the GC.
9. Restart the DSQ II.
10. Once the DSQ II has pumped down and is operational, check for vacuum leaks by
reading the convectron gauge and ion gauge pressures and running an automatic tune.
1
Transfer Line Gas Fitting
4
Transfer Line (PN 119640-1050)
2
Three Screws Around the Transfer
Line
5
O-ring (not shown)
3
Transfer Line Cable
1
2
4
3
5
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...