3 • SP Sample Preparer Description
Doc. 065250-06 03/2020
45
3.7
Analog Inputs
The main SP electronics board includes four analog inputs. An analog input cable
(P/N 070171) is provided in the SP Ship Kit (SP1, P/N 069085; SP2, P/N 069086)
for connection of an analog input to an external device. If you need more than one
analog input, order additional cables separately.
The input voltage range for each input is 0 to 10 V (maximum). The sampling rate
is 3 Hz and the resolution is 24 bits.
To connect the analog input:
1. Turn off the SP power.
2. Disconnect the power cord.
3. Follow the instructions in
to access the main SP electronics
board.
4. Locate the analog input connectors in the lower-left corner of the main SP
Figure 3-26. Main SP Electronics Board: Analog Input Connectors
1
2
3
4
Analog Input 4
Jumper
Analog
Input
Connectors
Содержание Dionex
Страница 1: ...Dionex Integral Process Analytical Systems Operator s Manual Document No 065250 Revision 06 March 2020...
Страница 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Страница 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Страница 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Страница 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Страница 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Страница 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Страница 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Страница 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Страница 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Страница 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Страница 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...