UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Page
60
Operating Instructions
4.2.2 Routing Capillaries
When routing capillaries in the module or to other system modules, observe the following
general information:
Pump Module
•
Two passages in the enclosure bottom of the pump module (
→
Fig. 4, page 27) facilitate
routing the capillaries to the modules that are located below the pump module and ensure
the shortest possible connection.
•
Two clips are installed in the pump module, for example, to secure the solvent lines to
the interior front panel of the pump.
Fig. 21: Clips in the pump module
Column Compartment
•
Capillaries to be routed to the outside
on the sides
of the column chamber
have to be
routed through the guides provided on the left and right of the column chamber.
Fig. 22: Capillary guides (here on the right of the column chamber)
•
Capillaries to be routed to the outside
on the top or bottom
of the column chamber can be
routed out of the interior at any position between the top or bottom of the enclosure and
the front panel door to make sure that the connections are as short as possible.
Preinstalled clips
Содержание Dionex UltiMate 3000 Series
Страница 2: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Operating Instructions ...
Страница 8: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page vi Operating Instructions ...
Страница 24: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 16 Operating Instructions ...
Страница 62: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 54 Operating Instructions ...
Страница 92: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 84 Operating Instructions ...
Страница 246: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 238 Operating Instructions ...
Страница 262: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 254 Operating Instructions ...