Doc. 031880-04 1/12
3. Set the
Download/Operation
button on the Dionex RFC rear panel (see
Download
(in) position.
4. Turn on the Dionex RFC power.
5. Insert the diskette in the computer’s drive. Display the diskette
contents, using My Computer or Windows
®
Explorer and
double-click the RFC icon
. A dialog box appears with a list of
serial ports.
6. Select the computer serial port and click
OK
. A Moduleware
download dialog box appears.
7. Click
OK
to start the download.
When complete, the Dionex RFC screen displays the new
Moduleware version number.
8. Set the
Download/Operation
button on the Dionex RFC rear panel to
the
Operation
(out) position.
6.2
Calibrating the Dionex RFC
When to Calibrate
Calibration is done at the factory before shipping. Thereafter, the Dionex
RFC should be calibrated if a new electronics board is added or before a
validation procedure.
How to Calibrate
1. Press the
or
button to display the calibration screen (see
2. Press
to select
YES
.
3. Press
ENTER
.
The message “Calibration in Progress” is displayed. When calibration
is complete, the display returns to the default status screen.
Figure 42. Calibration Screen