Shutting Down the Dionex RFC
Doc. 031880-04 1/12
37
Setting the Ion Count Percentage
1. Press
or
to display the EGC ion count screen (see Figure 41).
2. Press
ENTER
.
The
ION COUNT =
value on the screen is highlighted.
3. Press
or
until the desired count is displayed. If this is a new
cartridge, select 100%. If you are reinstalling a used cartridge, select
the value that was displayed before the cartridge was removed.
4. Press
ENTER
.
4.
Shutting Down the Dionex RFC
Figure 41. EGC Ion Count Screen
DX-120 Shutdown
Front panel control
Turn off the pump.
Software control
Turn off the pump from the software.
DX-320/DX-320J Shutdown
Front panel control
Set flow to 0
or
Turn off the pump motor.
Software control
Write a program or method to set the flow to 0.
This disables the TTL and shuts off current to
the suppressor.
DX-500/DX-600 Shutdown
Front panel control
Set the flow to 0
or
Turn off the pump motor.
Software control
Write a program or method to set the flow to 0.
This disables the TTL and shuts off current to
the suppressor.