Dionex ICS-5000 Operator’s Manual
218
Doc. 065342-06 12/11
Repeat two or three times. If the baseline does not improve, check the
other causes of baseline instability described in this section.
NOTE
To prevent air from becoming trapped in the cell in the
future, increase the backpressure on the cell by con-
necting backpressure tubing to the cell outlet. The
backpressure limit for the ED
cell is 690 kPa
(100 psi). Do not exceed this limit.
3. The water used to prepare the eluent may contain trace contaminants.
Remake the eluent, using ASTM Type I (18 megohm-cm) filtered and
deionized water that meets the specifications listed in
.
8.3
Poor Retention Time Reproducibility
•
Liquid leaks
1. Check for leaks from the piston seals. Replace the piston seal on any
head with a leak (see
2. Check for leaks throughout the remainder of the system, including the
check valves, injection valve, and columns. Tighten or replace fittings
as needed.
•
DP/SP not primed
•
Liquid lines incompletely flushed after an eluent change
Attach a 10 cc syringe (P/N 054578) to the waste port on the priming
valve. Open the priming valve (turn the knob one-half turn
counterclockwise). Draw at least 20 mL of the new eluent through the
liquid lines before operation.
•
Inoperative eluent proportioning valve (gradient pump only)
The proportioning valve assembly should be replaced. Contact Technical
Support for Dionex products for assistance.
Do not block the end of the cell outlet tubing for longer than the
recommended 2 to 3 seconds. Doing so creates high backpressure,
which can break the pH-Ag/AgCl reference electrode glass membrane.
Содержание Dionex ICS-5000
Страница 29: ...1 Introduction Doc 065342 06 12 11 13 Figure 1 3 Chromeleon 6 8 Panel Tabset Dionex ICS 5000 Home Panel Shown ...
Страница 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Страница 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Страница 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Страница 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Страница 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Страница 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Страница 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Страница 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Страница 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Страница 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Страница 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Страница 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Страница 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Страница 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Страница 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Страница 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Страница 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Страница 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Страница 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Страница 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Страница 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...