3 • System Setup
Doc. 065447-02 12/14
33
3.4.1
Installing the IC Cube Component Cartridges
The Dionex IC Cube houses three removable component cartridges and one
column tray for running capillary IC applications. For reference,
shows an IC Cube after these items have been installed.
The required cartridges depend on the detection type (see
Capillary
IC
Figure 3-6. Example Dionex IC Cube with Cartridges and Column Tray
Installed
Detection Type
Required Cartridges
Conductivity
•
EG degas (typically installed at the factory)
•
CES
•
CRD (an option for anion applications)
•
CRD bypass (required if a CRD cartridge is not
installed)
•
Column tray
Electrochemical
a
a. If you are setting up the system for electrochemical detection and the EG degas
cartridge is already installed, no additional cartridges are required. Leave the sup-
pressor bypass and CRD bypass cartridges in place and go on to
.
•
EG degas (typically installed at the factory)
•
Suppressor bypass
•
CRD bypass
•
Column tray
Table 3-1. Dionex IC Cube Cartridge Requirements
EG Degas Cartridge
CRD Cartridge or
CRD Bypass Cartridge
Suppressor Cartridge or
Suppressor Bypass
Cartridge
Column Tray