B • Installing DC Options
Doc. 065447-02 12/14
129
5. Remove the protective film from the alignment pins on the cell body (see
6. If you are installing the ED cell in a capillary IC system, use an open-end
wrench to remove the titanium inlet line from the cell (see
7. Rinse the surface of the cell with ASTM Type I (18 megohm-cm) filtered
and deionized water.
8. Clean the polished surface of the cell with a damp clean, lint-free tissue.
9. Install the working electrode. For a disposable working electrode, see
. For a conventional working electrode, see
.
Figure B-3. Removing the Cell Protective Film
Figure B-4. Capillary IC Only: Removing the Titanium Inlet Line
Remove the
cell protective
film
Capillary
IC
Capillary IC Only:
Remove the
titanium inlet line