ICS-2000 Ion Chromatography System
4-12
Doc. 031857-03 4/06
This error message appears when you click the
Shutdown
button on the
Chromeleon Control panel. The message also appears anytime you turn off
the pump flow while the suppressor is on. The suppressor is automatically
turned off to prevent damage to the suppressor.
To troubleshoot:
•
If the pump stopped unexpectedly, see
troubleshooting information.
4.2
Liquid Leaks
•
Leaking fitting
Locate the source of the leak. Tighten or, if necessary, replace the liquid line
connection (see
Installation of Dionex Liquid Line
Fittings
(Document No. 031432) for tightening requirements. The manual is
on the Dionex Reference Library CD-ROM (P/N 053891).
•
Broken liquid line
Replace the line and fittings with the same length and internal diameter tubing
(see
•
Blocked or improperly installed line
Make sure the lines are not crimped or otherwise blocked. Also, if the blocked
line is a waste line, make sure it is not elevated at any point after it exits the
ICS-2000. If a line is blocked, replace it (see
•
Loose pump check valve
1. Make sure the check valves are firmly seated in the pump head. If they are
not, tighten them carefully with an open-end wrench just until the leak
stops.
2. If the leak persists, replace the check valve (see
•
Leaking seal wash port
Leaking from the seal wash port indicates a leaking piston seal. Replace the
seal (see
SUPPRESSOR STOPPED FOR ZERO FLOW RATE
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...