D • TTL and Relay Control
Doc. 031857-03 4/06
D-11
Create an AS or AS50 (USB) Method and Schedule
is an example of AS or AS50 (USB) method timed events that can be
used with the TTL/Relay connections shown in
.
Timed Event Notes:
•
At time INIT, the injection valve is set to Load.
•
At time 0.00, the injection valve is set to Inject, an autozero is performed
(TTL2=1), the data acquisition mark is sent (RLY1=1), and the pump is
turned on (RLY2=1).
•
At time 0.5, TTL2 and RLY1 are switched back to 0 to be ready for the next
run. RLY2 remains at 1 to keep the pump on.
•
TTL1 does not need to be set in the timed events because the Load and Inject
commands control the ICS-2000 injection valve automatically (see
AS or AS50 (USB) Control of the Injection Valve” on page D-9
To run the above method, set up an AS or AS50 (USB) schedule. The example in
includes a four-point calibration (Lines 1 – 4) followed by two
samples (Lines 5 and 6).
Figure D-11. Example AS or AS50 (USB) Method for Stand-Alone Operation
Figure D-12. Example AS or AS50 (USB) Schedule for Stand-Alone Operation
Help Prompt
METHOD:
TIMED EVENTS
1
1
1
TIME
INIT LOAD
INJECT
0.00
VALVE
TTL1 TTL2 RLY1 RLY2
1
CSV
0
0
1
0.50
Help Prompt
SAVE TO
RUN
SCHEDULE
25.0
5.0
1
1
1
4
3
5
5
1
6
6
1
1
MISSING VIAL ACTION:
INJ/
INJ
VIAL
METHOD
VOL(uL)
VIAL#
LINE
END
START
EDIT
SKIP
1
1
1
1-4
5
6
-
-
-
-
25.0
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...