D • TTL and Relay Control
Doc. 031857-03 4/06
D-3
c.
Connect the wires from the ICS-2000 connector plug to the TTL or relay
connector pins on the other module(s). Additional connector plugs are
provided with other Dionex modules.
NOTE Check the polarity of each connection. Connect signal
wires to signal (+) pins and ground wires to ground (-)
pins.
3. If you connected a TTL input, verify that the correct function is assigned to
the input and that the correct input control type is selected. Select different
settings if necessary. Input functions and control types are assigned from
either the
INPUT/OUTPUT
touch screen page or Chromeleon (see
D.1.1
Selecting TTL Input Functions and Control Types
To select TTL input functions and the control type from Chromeleon,
open the Chromeleon Server Configuration program and double-click the
ICS-2000 icon under the timebase. Select the
TTL Inputs
tab (see
Figure D-3. ICS-2000 Server Configuration Properties: TTL Inputs
Содержание Dionex ICS-2000