Setting Up the Eluent Generator
Doc. 031857-03 4/06
B-45
9.
Slide the EluGen cartridge down
into the holder until it stops.
Make sure the cable and
backpressure coil remain out of
the holder (see
NOTE To avoid crimping the
lines inside the holder, do
not twist the cartridge
when you slide it into the
holder. The lines are
designed to coil inside the
holder.
Figure B-41. Installing the EluGen Cartridge in the Holder
10. Remove the plug from
the
EGC-1
connector on
top of the ICS-2000.
11. Orient the cartridge’s
blue 4-pin electrical
cable connector as
shown in
.
Push it onto the
EGC-1
connector on the ICS-
2000. Twist the ring on
the cable connector
fingertight to secure it.
Figure B-42. Connecting the EluGen Cartridge Cable