ICS-2000 Ion Chromatography System
B-30
Doc. 031857-03 4/06
B.12 Setting the Eluent Level
After filling the reservoir, open the touch screen
HOME
page and enter the volume
of liquid in the reservoir in the
Eluent Level
field (see
), or enter the
volume in the
Eluent Bottle
field on the Chromeleon
Control panel (see
). The ICS-2000 determines the eluent usage, by monitoring the flow
rate and the length of time the pump is on, and updates the field as the eluent is
used up. A warning is displayed if the level falls below 200 mL. The warning is
repeated at 100 mL and at 0 mL.
Go on to
In order for the eluent level displayed on the screen to be accurate,
the level must be entered by the user when the reservoir is filled. The
ICS-2000 does not automatically detect when the reservoir is filled,
nor when it is empty.
Figure B-25. Home Page: Setting the Eluent Level
Figure B-26. Chromeleon Control Panel: Setting the Eluent Level
Enter the volume of
liquid in the reservoir
Note: The Eluent Level
field is updated as the
liquid is used up.
Enter the volume of
liquid in the reservoir
Note: The Eluent Bottle
field and slider gauge
are updated as the
liquid is used up.
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...