Setting Up the Eluent Reservoir
Doc. 031857-03 4/06
B-29
B.11 Setting Up the Eluent Reservoir
1. Rinse the eluent reservoir.
2. Fill the reservoir with ASTM filtered, Type I (18-megohm) deionized water.
3. Place the reservoir on top of the ICS-2000.
4. Install the cap and handtighten.
NOTE The ICS-2000 does not require pressurized reservoirs.
However, if eluent is manually degassed or is sensitive to
contamination, Dionex recommends pressurizing the
reservoir with helium or nitrogen. See
for
instructions.
5. Go on to