ICS-2000 Ion Chromatography System
5-38
Doc. 031857-03 4/06
3. Replace the two fuses in the holder with new IEC 127 fast-blow fuses rated
3.15 amps (P/N 954745). Dionex recommends always replacing both fuses.
4. Reinsert the fuse holder into its compartment and push in until the tab clicks
in place.
5. Reconnect the main power cord and turn on the power.
5.17 Replacing an EluGen Cartridge
The procedure for replacing an EluGen cartridge differs, depending on the type of
cartridge:
•
To replace a KOH, LiOH, MSA, or NaOH EluGen cartridge, go to
•
To replace a K
2
CO
3
cartridge, go to
The EluGen cartridge contains one of the following: a corrosive base
(KOH, LiOH, or NaOH), or a corrosive acid (MSA). Wear protective
eyewear and gloves when handling the cartridge.
La cartouche d'EluGen contient un de ce qui suit: une base corrosive
(KOH, LiOH, ou NaOH), ou un acide corrosif (MSA). Porter des
lunettes et des gants protectives en manipulant la cartouche.
Die EluGen-Kassette enthält eine korrodierende Base (KOH, LiOH
oder NaOH), oder eine korrodierende Säure (MSA). Tragen Sie daher
beim Umgang mit Kassette eine Schutzbrille und Handschuhe.
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...