5 • Service
Doc. 065289-03 10/12
109
a.
Open the Wellness Panel (see
) and click
Calibration
Details
.
b. Under
Column Heater Calibration Values
, enter the
Column Heater
Offset
and
Column Heater Slope
values recorded in
.
c.
In the field next to the
Download
button, verify that
Current
is selected
and then click
Download
to download the new calibration values to the
Dionex ICS-1100.
d. To ensure the new calibration values are recorded in the Dionex ICS-1100
memory, do not turn off the Dionex ICS-1100 power for at least 1 minute
after downloading the values.
5.13 Replacing the Column Heater Heat Exchanger
1. Turn off the Dionex ICS-1100
power.
2. Open the front door of the
Dionex ICS-1100.
3. Unscrew the two thumbscrews
on the column heater cover
(they remain attached to the
cover).
4. Pull the heater cover straight
out to remove it.
5. Unscrew the six thumbscrews
) and remove the
plate.
6. Disconnect the heat exchanger
inlet line from port C on the
injection valve.
Figure 5-17. Dionex ICS-1100 Column Heater Cover
Removed
Heat
Exchanger
Screws (3)
Thumbscrews
(6)
Содержание Dionex ICS-1100
Страница 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Страница 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Страница 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Страница 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Страница 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Страница 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Страница 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Страница 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...