Maintenance
86
|
Dionex IC Pure system
Thermo Scientific
Chapter 13
Changing the ultrafilter
6
a.
Switch off the unit, remove the disinfection adapter by
pressing on the quick-connectors and install the new
ultrapure cartridge (see “Changing the ultrapure cartridge”
on page 79).
b.
If an optional pretreatment cartridge is installed, remove the
cartridge tube jumper by pressing on the quick-connectors at
the connections for the pretreatment cartridge and then
install a new pretreatment cartridge (see “Replacing the
pretreament cartridge (optional)” on page 80).
7
Switch the system on.
Step
Action
Figure
Tube jumper
Quick-
connectors
b.)
Adapter for disinfection
Quick-connectors
a.)
Note
You will need the following tools for replacing the ultra-filter: Open-end wrench, size 17,
Phillips screwdriver and Teflon strip
Step
Action
Figure
1
Switch the system off and unplug from power supply
Содержание dionex ic pure water
Страница 10: ......
Страница 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Страница 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Страница 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Страница 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Страница 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Страница 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Страница 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Страница 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Страница 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Страница 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Страница 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Страница 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Страница 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Страница 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Страница 127: ...Thermo Scientific IC Pure 117 ...