Maintenance
|
Chapter 13
Autoclaving the 0.2
μ
m final filter
Note
The lifetime of the sterile filter can be prolonged by sterilizing it. Proceed as follows to
autoclave the sterile filter.
Step
Action
Figure
1
Rotate the used sterile filter in a clockwise direc-
tion to remove it.
2
Use an autoclave to sterilize the sterile filter in a
steam autoclave at 121 °C for 30 minutes. You can
repeat the sterilization process for the sterile filter
up to 5 times.
When sterilization is completed, screw the sterile
filter back into the 1/4“ thread on the sampling tap
(see “Replacing the 0.2 µm final filter” on
page 92).
1/4“ thread
0.2 µm final filter
Note
If you wish to remove purified water and the flow
rate is too low, this is an indication that the sterile filter
is clogged. In this case, read the description in section
“Troubleshooting” on page 99, or replace the sterile
filter with a new one.
Содержание dionex ic pure water
Страница 10: ......
Страница 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Страница 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Страница 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Страница 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Страница 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Страница 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Страница 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Страница 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Страница 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Страница 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Страница 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Страница 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Страница 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Страница 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Страница 127: ...Thermo Scientific IC Pure 117 ...