Dionex Easion Operator’s Manual
56
Doc. 155028-01 08/20
4.9
Nonreproducible Peak Height or Retention Time
•
Column overloading
Dilute the sample (see
•
Liquid leaks
Locate and eliminate the leaks (see
•
Incomplete or imprecise filling of the sample loop
1. Fill the sample loop until excess sample exits the waste line.
2. Inspect the 1 cc syringe (P/N 016388) and replace if damaged.
•
Pump not primed properly
4.10 Abnormal Retention Time or Selectivity
•
Contaminated or incorrect eluent
Remake the eluent, using concentrated eluent and ASTM Type I
(18.2 megohm-cm) filtered and deionized water (see
•
Contaminated or degraded sample
Take appropriate precautions when preparing and storing samples to prevent
contamination and degradation (see
•
Contaminated column
1. Clean the column as instructed in the column manual.
2. If cleaning is unsuccessful, replace the column.
4.11 No Detector Response
•
Cell not properly installed
Check that the cell is plugged into the component mounting panel and the cell
housing is screwed down so that the bottom of the housing is flush against the
sheet metal panel (see
and
Содержание Dionex Easion
Страница 1: ...Dionex Easion Ion Chromatography System Operator s Manual Document No 155028 Revision 01 August 2020 ...
Страница 34: ...Dionex Easion Operator s Manual 26 Doc 155028 01 08 20 ...
Страница 68: ...Dionex Easion Operator s Manual 60 Doc 155028 01 08 20 ...
Страница 76: ...Dionex Easion Operator s Manual 68 Doc 155028 01 08 20 Figure 5 2 Dionex Easion Flow Schematic ...
Страница 110: ...Dionex Easion Operator s Manual 102 Doc 155028 01 08 20 ...
Страница 118: ...Dionex Easion Operator s Manual 110 Doc 155028 01 08 20 ...