3 • Operation and Maintenance
Doc. 155028-01 08/20
39
3.11.2 Manual Sample Processing
To manually process a sample, select operating parameters and
commands from the Chromeleon ePanel. Commands are executed as soon
as they are entered.
Summary of Manual Sample Processing
1. Complete the instructions in
prepare the Dionex Easion for operation and to prepare the sample for
processing.
2. Load the sample into the injection valve sample loop with either the
3. Click
Autozero
on the Chromeleon ePanel.
4. Switch the injection valve to the Inject position with either the
The signal plot is displayed on the Chromeleon
ePanel.
5. Monitor the chromatogram. When sample data has been collected,
click
Monitor Baseline
on the Instrument toolbar above the
Chromeleon ePanel Set.
Saving Manual Data
Data from manual processing is saved in the
manual
sequence under the
instrument folder in the local data vault.
To save the data from a manual run:
1. Select the
manual
folder and select
File > Save As
.
2. Enter a new name for the sequence.
3. Select the
Save raw data
check box.
4. Click
Save
.
Содержание Dionex Easion
Страница 1: ...Dionex Easion Ion Chromatography System Operator s Manual Document No 155028 Revision 01 August 2020 ...
Страница 34: ...Dionex Easion Operator s Manual 26 Doc 155028 01 08 20 ...
Страница 68: ...Dionex Easion Operator s Manual 60 Doc 155028 01 08 20 ...
Страница 76: ...Dionex Easion Operator s Manual 68 Doc 155028 01 08 20 Figure 5 2 Dionex Easion Flow Schematic ...
Страница 110: ...Dionex Easion Operator s Manual 102 Doc 155028 01 08 20 ...
Страница 118: ...Dionex Easion Operator s Manual 110 Doc 155028 01 08 20 ...