Dionex AS-DV Operator’s Manual
68
Doc. 065259-05 10/12
•
Sample is viscous
Reduce the delivery speed by 20% or more in the Chromeleon instrument
method or program (see
4.8
Autosampler Jammed
•
Vial carousel is jammed
1. Press the
Carousel Release
button to try to home the carousel.
2. Remove the carousel (see
) and check for foreign objects.
•
Sampling tip jams as it is being withdrawn
1. Check for excessive backpressure (see
2. Check the sampling tip for possible obstruction:
a. Turn off the power to the Dionex AS-DV.
b. Manually raise the sampling head (see
c. Check for blockages.
Содержание Dionex AS-DV
Страница 1: ...Dionex AS DV Autosampler Operator s Manual Document No 065259 Revision 05 October 2012 ...
Страница 8: ...Dionex AS DV Operator s Manual vi Doc 065259 05 10 12 ...
Страница 30: ...Dionex AS DV Operator s Manual 22 Doc 065259 05 10 12 ...
Страница 126: ...Dionex AS DV Operator s Manual 116 Doc 065259 05 10 12 954745 Fuse IEC 127 rated 3 15 A 1 Part Number Item Quantity ...