Operation
Detector Screens
4
-
36
DensityPRO User Manual
Thermo Scientific
Detector
Configuration
The detector configuration parameters on the Detector Setup tab provide the user with
the ability to view and configure high voltage power supply parameters for the
detector. The gauge has the ability to read and write any of these parameters as
requested by the user. When the system is shut down, the last stable high voltage
control information will be used at the next startup.
Each gauge can utilize up to two high voltage power supplies.
1.
The Force Hi Voltage #0 Hold Value dropdown provides the user with the option
to place the detected high voltage power supplies in hold mode.
a.
If Yes is selected, the system will hold the high voltage at the value designated
in the Hi Voltage #0 Value textbox.
b.
If No is selected, the system will use live system data as the Hi Voltage #0
Value.
2.
The following fields are intended for display only, however, the user has the ability
to modify the values in these fields.
Last HV Control Time Constant
Last CPLD Window Count
Last Board Temp ºC
Last Hi Voltage #0 Control
Last Hi Voltage #0 Setup
Min CPLD Channel 1 Window Count
Warning:
Making changes to these parameters may cause the system to
become unstable.
Содержание DensityPRO NAI Series
Страница 2: ......
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 14: ...xiv DensityPRO User Manual Thermo Scientific This page intentionally left blank ...
Страница 22: ...A xxii DensityPRO User Manual Thermo Scientific Quick Setup This page intentionally left blank ...
Страница 124: ...Operation Application Screens 4 72 DensityPRO User Manual Thermo Scientific Figure 4 39 High and Low Alarms ...
Страница 134: ......
Страница 158: ......
Страница 195: ......